www.kraft-butzke.de

Herzlich Willkommen auf www.kraft-butzke.de
             Firma Kraft & Butzke Leipzig

Vacuscope  II    -   simply   professional

VACUSCOPE II is a handy meter for checking pulsators and measuring the vacuum range in milking installations. Among, the  outstanding product features wich characterize Vacuscope are:

  - Convenient use due to it´s compact design
 - Menu guidance
 - During testing, the Vacuscope visually displays the vacuum course  
 - LCD Display (128x64pixel) with illumination
  - Interface connection RS 232 for external sensors, Personal Computer (PC), printer and air-flow meter
 - All readings can be saved in the memory
 - Changing the measuring units (kPa or in.HG)

All available programs are shown in the main menu:.

 

Pulsation

Check the milking installations´ pulsation parameters according to ISO 5707:
 

                

  - Simultaneous measurement at two channels
  - Memory capacity for 100 readings
  - Displaying the pulse phases in [ms] and [%]

Fluctuation

 

 - Measurement of vacuum and pressure
 - The measuring interval is 10 ms
 - different graphical variety

  
 



Gauge

 - Measurement of vacuum, vacuum differences and pressure
- The measuring interval is 100 ms

 ISO Test

  - Complete Installation test according to ISO 
     6690:1984 and according to ISO 6690:1996 
    

- Saving 4 test reports
- Automatically calculation of all values
- Automatically calculation of all limits
- Automatically issue of records
- Displaying and printing the test report or parts

Technical Data:

 

Measuring range:

Vacuum     0-80 kPa
Pressure     0-80 kPa

Precision:

Vacuum: +/- 0,6kPa
             +/- 0,3 kPa(typ.)
Pressure: +/- 0,8kPa
External sensors +/- 0,6kPa
RS 232 for external

Interface connection :

sensors, PC and printer

Screen:

LCD-Display (128x64 pixel) with automaticallycontrolled illumination

Memory capacity:

100 readings with 320 points
each4 test reports
Air admission of 16 clusters

Power supply:

4xR6 Alkaline batteries;
1,5V each

Tests:

EN 50081-1, EN 50082-1